Friday, July 12, 2013

Product Catalogue of Implant System

The following catalogue contains, apart from elements overview, basic
information about products made by Dyna Dental Engineering. It is intended to give
the user a clear overview of our products, extra general technical data, as well as
practical tips concerning clinical and technical procedures. It is, however, not the full
product description and it may serve only as a guide through our products.
Dyna Implant SystemBecause of the fact that all Dyna implants are produced in four different
diameters there exist four different lines of implants (Ø3.0, Ø3.6, Ø4.0, Ø5.0mm) and
corresponding with them prosthetic elements. To facilitate the dentist and the
technician recognition of a particular line all elements are colour coded.
The following colours have been chosen for particular implant diameters:
Ø 3.0 White (no colour, colour of titanium)
Ø 3.6 Magenta
Ø 4.0 Blue
Ø 5.0 Yellow
The colour is achieved in the process of titanium
oxidation and therefore has the same properties as
titanium in relation to biocompatibility and tissue
response.
All products are coloured in such places that they do
not influence the final aesthetic result, but at the
same time offer easy recognition of diameter during
all phases of surgery and prosthetic construction

-----------------------------------------------------
Planning aids
X-ray tracing
To facilitate implant choice in a particular clinical situation the X-ray tracing can be used. Thin
transparent plastic foil with all fixtures contours in 1:1 and 1:3 (for OPG!) scale put over the X-ray
photo allows to determine implant type and right position.
Please notice:
Because of the fact that X-ray photos are rarely exact it is important to
remember that this method is not entirely precise. In more demanding
situations we recommend using CT scans.
-----------------------------------------------------

Implants
Dyna implants are two-stage, endosseous, push-in dental
fixtures. They are sterilized in the process of gamma irradiation
and supplied in sterile double peel-pouch packaging.
The current concept of Dyna is based on the full, reverse Edisonscrew
implants covered fully (without polished collar) with
hydroxylapatite. The essential features of this design are:
-fast osteointegration
-good physiological loadability
-direct bone apposition
-simple and time-saving implantation technique
-favourable load distribution.
-good results in bad quality bone
Coating titanium implants with HA, besides excellent
biocompatibility, favours direct bone growth to the implant
surface. In contrast to the other non-HA implants the process of
the bone apposition occurs not only from the bone side but from
the implant surface as well.
The implants are already provided with a titanium
cover screw (height 0,4mm)and suspended on a plastic, colourcoded
holder, facilitating manipulation during the surgery.
The implants are available in diameters 3,0; 3,6; 4,0; 5,0
and lengths 6 (only for maxillo-facial purposes) 8,10,13,15mm
and 17mm(for Ø3, Ø3,6).

-----------------------------------------------------
Planning aids
Implant X-ray tracing 5472
Implants HA Coated Ø3.0






Implants HA Coated Ø3.6






Implants HA Coated Ø4.0
 




Implants HA Coated Ø5.0

Closing screws

-----------------------------------------------------

Drills
Together with the system a special set of drills is supplied. All drills supplied by
Dyna Dental Engineering, produced for Dyna Implant System, are made of surgical stainless
steel and apart from the pilot drill internally irrigated.
Depending on the quality of the bone they can be used for maximum of 20 implantpreparations.
All drills have markings corresponding with the several implant lengths.
They are, however, positioned slightly higher than the real implant length in order to prepare
a bone bed ready to accommodate an implant together with the covering screw (0,4mm). In
this way the implant shoulder does not protrude over the bone ridge and there is some space
for possible initial bone resorption.
Please notice:
Always prepare the implant site so that the desired length marking is covered slightly under the bone
ridge.
Dyna drills are intended to be used with the surgical contra angle with internal cooling
system, and a special surgical unit.

Pilot drills
are used to make the initial drilling. With these drills the depth of the drilling and insert
angulation are defined. (see also 6.3 and 7)
In situations where conditions do not allow for further drilling the initial holes can be left
empty and the new more favourable place for the implant found
.







Spade drills

are intended to be used for final bone bed preparation. Five different types of internally
cooled drills allow preparation to the desired diameter.
Please Notice:
It is obligatory to follow the sequence of widening of the preparation (see chapter 7).
All drills have special markings corresponding with the implants’ lengths and facilitating orientation
in the depth of drilling.











Special drills

are to be used to prepare a place for initial drilling, remove sharp bone edges or slightly
change the primary insert angulation. Lindemann fraise may additionally be used instead of
a pilot drill.

-----------------------------------------------------
Surgical instruments
All surgical instruments are made of surgical stainless steel or titanium ELI 5 (TiAlV).
In order to obtain parallel preparations, the parallel/depth instrument can be placed in the
implant socket, thus visualizing the direction of the preparation. The instrument is colour
coded and serves also for measuring depth and diameter of the preparation.
To pull out implants that have been accidentally placed too deep Dyna implant puller
should be used.
In order to remove the problematic implant the covering screw must be unscrewed and
replaced by proper implant puller (corresponding diameter; colour coding!). With the
implant puller, turning clock wise, it is possible to carefully remove the implant without
touching or contaminating it.
For screwing or unscrewing elements during the surgery surgical screwdrivers should be
used.
A hole in the screwdriver’s handle serve as an extra protection against aspiration ( after
fixing it on e.g. dental floss)
Auxiliaries
Locking drill extender
The locking drill extender is a drill holder used to extend the length of drills in demanding situations.
Cleaning wire
To clean the internal cooling system of spade drills use a cleaning wire.
To organize in an easy way all Dyna surgical instruments and drills use sterilizable cassette for
surgical instruments and drills.
Please note:
This product is manufactured from coloured anodised aluminium. The colour of the product may be
damaged by the use of some chemicals.
Remember:
-do not use dish washer
-do not use ultrasonic bath
-be careful when using disinfecting liquids
-do not use chemicals too alkaline or too acid; pH between 4 and 8 is safe
-for sterilizing follow the following process: autoclave (134o), chemiclave (132 o) or dry heat
(180o). For cycles and time please refer to the manufacturer’s instructions.
----------------------------------------------------

Dyna Parallel/Depth Instrument

Dyna Pull out instrument

Auxiliaries























---------------------------------------------------------------------

Drivers
In order to screw final prosthetic abutments to 30 or 35Ncm* or unscrew them Dyna torque
wrench prosthetic screwdrivers should be used.
S - Drivers were designed to facilitate handling of Dyna products. This line of screwdrivers, besides
better mechanical properties, has additionally a special connection for the Straumann® Torque
Wrench. In this way both Dyna and Straumann® instruments may be used interchangeably. The new
connection part facilitates grip of the instrument as well.
Dyna Hex Driver T.W.-S serves for screwing and unscrewing all elements with hex opening. It can
be used with Dyna or Straumann Torque Wrench*.
Dyna Single Slot Driver T.W.-S serves for screwing and unscrewing all elements with slot opening.
It can be used with Dyna or Straumann Torque Wrench*.
Castellated driver TW-S serves for screwing and unscrewing “castellated” abutments. It can be
used with either Dyna or Straumann® Torque Wrench*.
Notched driver serves for screwing fixed GN and memory GN abutments and adjusting the
angulation of the memory abutments.
Torque Wrenches
Torqe wrench Is a special instrument intended to screw all final abutments with 30 or 35Ncm*
torque. This torque prevents abutments, subjected to physiological bite forces, from unscrewing.
(see also instructions for use delivered with the Dyna Torque Wrench)
* Apply 35 Ncm indication for all abutments placed directly on implant level
Apply 30 Ncm indication for all abutments placed on extension level
Other Instruments
Grip Device
To make the grip of T.W. instruments easier a special holder can be applied.

-----------------------------------------------------------------------
Healing Abutments
Healing abutments are a special type of titanium screws used in the second stage surgery to
heal and form the gingiva. They have height markings on 2mm, 4mm and 6mm (the top)
allowing for gingiva height determination after the healing period.
Depending on the type of prosthetic construction different healings should be used for
overdentures and crown and bridge works.
Shape difference was designed not only because of hygienic requirements but because of the
fact that for crowns and bridges special aesthetic aspects should be taken into consideration
(emergency profile).
The height of healing abutments is 6 mm (without the screw).
Extension Posts
These abutments are used to extend bar or ball abutments above their standard lengths. They should
be screwed with single slot screwdriver.





















------------------------------------------------------------------------------------

Dyna magnetic attachment consist of:
 magnet
 ferromagnetic part (medical abutment,
prefabricated Dyna Direct keeper, Dyna alloy for
casted keepers)
Magnetic abutments are specially designed abutments for use with Dyna magnets.
They are produced from PdCoPt alloy.
They should be screwed with using Dyna TW-S screwdriver and torque wrench.
Dyna Dental produces also magnetic abutments for other systems.











Magnets

Dyna magnets are permanent rare earth magnets supplied in two attraction forces
(Normal Strength, Extra Strong) and two heights(1,5mm/NSS/ and 2,5mm/NS, ES/).
To prevent any corrosion they are encapsulated with 316L stainless steal. For full
magnet description please see the Dyna Magnet Manual.
Magnet Bond
To fix the magnets into the dentures use Dyna Magnet Bond, a specially designed
resin material. Its features enable to achieve chemical connection between the
magnets and a denture acrylic. Therefore, no discolouration appears around the
magnets.
Dyna Magnet bond may be used additionally in all situation when perfect bonding of
acrylic to metal is necessary.
























------------------------------------------------------------------
Ball attachment consist of:
 Ball abutment (Ti)
 Matrix (Au/Pd/Pt)
 PVC ring
 Impression transfer
 Tin foil ring
 Lab analogue
To screw the abutment use a hex screwdriver TW-S and torque wrench.
Please notice
The ball abutment is supplied together with impression transfer. Other elements must be ordered separately
Ball abutment instruments.
To activate or deactivate (increase or decrease the grip force) ball attachment two separate
instruments are required :
-activator and
-deactivator.
During the activation lamellae of the matrix are symmetrically bent closer increasing the
retention whereas during deactivation it is pushed aside.
Please notice:
When activating make sure that there is no PVC ring around lamellae of the matrix


  ------------------------------------------------------------------
Dyna ball inserting instrument
Dyna ball space retainers (PVC rings) are specially designed for retaining space around the matrix
during waxing and finishing of a denture. They are supplied together with the matrix (placed around
the lamellae). It is recommended to leave them in place in ready prosthesis. In some situations,
however, they have to be removed (e.g. activation, rebasing). To put them easily back without
destroying the matrix one should use the Dyna Ball Inserting instrument.
It consist of two parts. To place the PVC ring around the matrix follow the next procedure
 separate the handle and the core
 put two pvc rings on the core (they can be
inserted only from one side)
 put the handle back on the core
 place the instrument inside the matrix
 gently push on the handle sliding the ring around lamellae.

Impression copings.
Impression coping is a special element used during impression taking. It was designed for open
tray technique. It transfers exactly the implant position into the model.
Please notice:
These impression copings are meant to be used when producing bar overdentures. In certain clinical
situations they can be also used for other constructions.

---------------------------------------------------------------------------------

Bar attachment consist of:
 Bar abutment (consist of two parts the titanium sleeve and the
dentist screw)
 Lab screw
 Lab analogue
 Castable plastic ring
The Dyna bar abutment allows the dentist to decide on the type of the bar in a particular case.
Cemented bar guarantees totally stress free construction that can be still retrievable. In this way
there are no destructive lateral forces transferred onto the implants. Dyna Dental does not supply
any pre-fab bar itself giving the operator free choice.
Please notice:
Bar Abutment is supplied together with the castable plastic ring and screw. Other elements have to be ordered
separately!


-------------------------------------------------------------------------------------
The Universal (cast to) Abutment.
This versatile abutment expands the prosthetic possibilities and may be used in almost all
clinical situations. Custom fabrication from the top of the implant is a great advantage in
those cases where optimal emergence profile and crown design is necessary. Prosthetic
correction of the implant divergence or unfavourable angulation improvement is easy to
realize. Both screw-type and cemented constructions can be produced with help of this
universal abutment.
After taking the impression it is the technician who waxes up the desired
abutment first, and than produces planned prosthetic construction. The
castable sleeve can be easily prepared and adjusted to every desirable shape.
Low metal collar allows for achieving perfect aesthetics (for diameter 4mm
only a fully castable plastic sleeve is available) and achieve optimal
contouring of soft tissue, for an even more natural looking restoration.
The abutment is available for Dyna implants diameter 3, 3.6 and 4 mm.
Material: POM and Gold Alloy (61.5%Au, 20,1%Pt, 17.5%Pd, 0,5%Ag),
melting interval 1340oC -1470oC,
thermal expansion coefficient (25-500oC) 13.0μm/m.K,
recommended solders – Orion 1120oC white
general indication:
1. For soldering with Elephant I 850oC, II 800oC, III 750oC or Pallas II 830oC or III 750oC
2. To cast against with precious alloys
3. To cast against both precious or non-precious alloys
4. To cast against with non-precious alloys
POM needs to be burned out during 20 minutes with a minimum temperature of 260 and
maximum of 300 °C. Heatingcyclus of the oven is recommended on a maximum of 30 °C
per minute. If burning out is done faster than recommended there is a risk of burning POM
causing porous casting.
Variations: The universal abutment is for the Dyna Implant system (not Octalock)
available in one form
- metal basis without octagon and plastic sleeve (soldering)
The universal abutment can be used for single or multiple unit (splint) crown and bridge
restorations or bar supported overdenture restorations. Highly aesthetic custom restoration
attached directly to the implant may be realized with this type of abutment.
Abutment Screw
Each abutment has its own abutment screw. Conical design guarantees optimal anti-loosening
properties.
Material: Titanium grade 5.
Laboratory Screw
Each abutment has its own laboratory screw.
Material: brass
------------------------------------------------------------
Abutments for fixed constructions.
These abutments are intended to be used for
cemented crowns and bridge constructions.
They have been designed according to gingiva
guiding idea. The outer transmucosal part of the
Dyna Fixed Abutment is highly polished whereas
the crown part is roughened. The abutments can be
shaped to realise highly aesthetical prostheses.
GG abutments have diameter 6mm for Ø3,6, and Ø4,0 mm implants and 7mm for Ø5mm implants.
GN abutments have Ø5mm for Ø3,6 and Ø4,0mm implants and 5,5mm for Ø5mm implants.
All abutments for Ø3.0 implants have one diameter 5.0mm.
These variations allows easier interdental placement and avoiding unfavourable emergency profile
and together with the possibility of adjusting to the gingiva height achieve a pleasing effect.
The GG shape has following advantages:
-possibility to hide crown margin under gingiva
-no pressure on papillae
-possibility to produce highly aesthetical prostheses

Dyna fixed abutment (0o)
These abutments may be used when implant positioning is parallel or almost
parallel. Different trans-gingival heights and types, GG and GN, allow to
achieve excellent aesthetic effects. GG abutments with bigger diameter are
provided to replace more “massive” teeth, whereas GN, with smaller
diameter, may replace lower incisors or premolars.
The total height and transgingival height are presented in the form of two
digits e.g. 2/5. The first means transgingival (polished) height whereas the
second total height.


-----------------------------------------------------------------------

Dyna memory abutment
This abutment can be used in situations where the positioning of the implants is not
favourable (angulation more than 6o). The outer transmucosal portion (cup) of the
memory abutment is made from highly polished titanium. This portion of the
abutment is the only part which, in clinical application, will contact the body tissue.
The abutment head (patrix) that is contained within the
transmucosal cup is made from nitinol (shape memory alloy: see
chapter 3.3)
A special shape-memory metal used in this abutment allows the
abutment head to be bent from 0o to 15o after spraying it with a
special Dyna cooling spray. GG shape of the transgingival part
allows for constructing highly aesthetical prostheses.
GG abutments have diameter 6mm for Ø 3,6, Ø 4,0 mm implants
and Ø 7mm for Ø 5mm implants. GN abutments have Ø 5mm
for Ø 3,6 and Ø 4,0mm implants and Ø 5,5mm for Ø 5mm
implants. All abutments for Ø 3.0 implants have one diameter
5.0mm
By using these variations unfavourable emergency profile can be
avoided. The possibility of adjusting to the gingiva height allows
to achieve pleasing effect.
The GG shape has the following advantages:
 possibility to hide crown margin under gingiva
 no pressure on papillae
 possibility to produce highly aesthetical prostheses
Cooling aids
Memory abutments can be cooled down in the mouth of the patient with help of the Memory
Cooling Aid. The aid resembles in its shape saliva ejector tip. To use it properly the long end
should be connected to the ejector whereas the wider opening of the ball part should be positioned
around the memory patrix. Please make sure that the patrix is placed totally inside the ball. If it is
not the case you can easily adjust it by cutting it to shape. Subsequently connect the cooling spray
with the smaller hole and cool the abutment for around 3-5 seconds. Wait a moment and with the
Dyna Position Instrument change the angulation.
Laboratory elements (fixed)
The laboratory analogue is intended to be used by the technician as an abutment analogue
in the working model. This model can only be poured in after previous positioning of the
analogue in the impression. See also 8.5 and 8.11
.
GG/GN Laboratory Analogues “fixed”
GG/GN Laboratory Analogues “memory”
Laboratory parts – implant analogue
Laboratory parts (bar)
Laboratory parts (ball)

==================================================================
Dyna Octalock
Because of the fact that Dyna Octalock implants are produced only in three
different diameters (3.6mm, 4.0mm and 5.0mm) and all abutments can be fitted on both
implant types, there is no need for special colour coding. Therefore, none of the Octalock
products is specially coloured.
Implants
Two-stage, endosseous, push-in dental fixtures.
Sterilized in the process of gamma irradiation and supplied in
sterile double peel-pouch packaging.
The concept of Octalock implants is based on the full, reverse
Edison screw implants covered, either fully or with 1mm
machined polished collar, with hydroxylapatite. The essential
features of this design are:
-fast osteointegration
-good physiological loadability
-direct bone apposition
-simple and time-saving implantation technique
-favourable load distribution.
-good results in bad quality bone
Coating titanium implants with HA, besides excellent
biocompatibility, favours direct bone growth to the implant
surface. In contrast to the other non-HA implants the process of
the bone apposition occurs not only from the bone side but from
the implant surface as well.
The implants are already provided with a titanium
cover screw and suspended on a plastic, colour-coded holder,
facilitating manipulation during the surgery.
The implants are available in diameters 3.6, 4.0 and 5.0
and lengths 8, 10, 11.5, 13 and 15mm.


Straight abutment
Ideal for all those situations where the implants are placed parallel or almost parallel.
The unique simplicity of this abutment makes all the connected procedures easy and
fast. The possibility of the abutment trimming gives a chance to get always the best
aesthetic result.
Both the single tooth reconstructions and bridges can be realized with this abutment.
The straight abutment can only be used with cemented crowns and bridges.
Material: Titanium grade 5.
Variations : Straight Abutment is available in different transgingival lengths
L2=2mm*, L3=3mm*, L4=4mm*. All abutments have the octagon. It is indicated for
single and multiple unit cemented restorations where no angulation correction is
necessary.
*depends on the implant diameter see product catalogue
Memory abutment.
The memory abutment employs the idea of using the memory
metal to change the angulation of the abutment’s head. Cooling it
with a special Dyna Coolspray enables easy adjustment of the
head to the particular clinical situation.
For more information please see Dyna Implant Manual.
Material: Titanium, Nitinol
Variations: Memory Abutment is available in two transgingival
lengths L2=2mm*, L3=3mm*.
*depends on the implant diameter see product catalogue
Temporary Abutment
The temporary Abutment is intended to be used after second stage surgery to provide
the patient with a temporary restoration. The titanium sleeve offers a textured surface
for better connection with the covering material, when constructing provisional
restorations on Dyna Octalock implants.
Materials: Titanium grade 1
Variations: Temporary Abutment is available in one form to be trimmed individually
either by dentist or by technician.



The Universal (cast to) Abutment.
This versatile abutment expands the prosthetic possibilities and may be used in almost all
clinical situations. Custom fabrication from the top of the implant is a great advantage in
those cases where optimal emergence profile and crown design is necessary. Prosthetic
correction of the implant divergence or unfavourable angulation improvement is easy to
realize. Both screw-type and cemented constructions can be produced with help of this
universal abutment.
After taking the impression it is the
technician who waxes up the desired
abutment first, and than produces planned
prosthetic construction. The castable sleeve
can be easily prepared and adjusted to every
desirable shape. Low metal collar allows
for achieving perfect aesthetics (it is also
possible to choose fully castable plastic
sleeve) and achieve optimal contouring of soft tissue, for an even more natural looking
restoration.
The abutment is available for single tooth reconstructions and cemented bridges with
octagon, and for screwed bridges without.
Material: POM and Gold Alloy (61.5%Au, 20,1%Pt, 17.5%Pd, 0,5%Ag),
melting interval 1340oC -1470oC,
thermal expansion coefficient (25-500oC) 13.0μm/m.K,
recommended solders – Orion 1120oC white
general indication:
5. For soldering with Elephant I 850oC, II 800oC, III 750oC or Pallas II 830oC or III 750oC
6. To cast against with precious alloys
7. To cast against both precious or non-precious alloys
8. To cast against with non-precious alloys
POM needs to be burned out during 20 minutes with a minimum temperature of 260 and
maximum of 300 °C. Heatingcyclus of the oven is recommended on a maximum of 30 °C
per minute. If burning out is done faster than recommended there is a risk of burning POM
causing porous casting.
Variations: The universal abutment is available in three forms
A. metal basis with octagon and plastic sleeve (soldering)
B. metal basis without octagon and plastic sleeve (soldering)
C. plastic abutment (to be casted totally by the technician)
The universal abutment can be used for single or multiple unit (splint) crown and bridge
restorations or bar supported overdenture restorations. Highly aesthetic custom restoration
attached directly to the implant may be realized with this type of abutment.
Fixation Screw
Universal fixation screw fits on all abutments types. Conical design guarantees optimal antiloosening
properties.
Material: Titanium grade 5.
Variations: Short fixation screw is available only in one form.

Ceramic\Titanium Abutment (trimmable)
The aesthetic abutment was developed in order to meet high aesthetic demands posed by
patients nowadays. The use of ceramic abutment besides excellent aesthetics creates
optimal biocompatibility, which is mandatory for a long-term success. Ideal mechanical
and biological properties of the Zirconium Oxide guarantee predictable and stable
results. Conical design of the abutment allows for individual trimming and therefore,
the best
adaptation to the individual mouth conditions.
*Example of the ceramic abutment after preparation.
The abutment can be trimmed to the desired form by the dental technician. The use of
ceramic material gives the prosthodontist a chance to adapt the colour of the restoration
to the natural situation.
The ceramic abutment is supplied in the trimmable form (see. catalogue) so that
both straight and angulated abutments can be easily realised. Such solution gives more
versatility and saves the need to have different abutment lines.
Material: Zirconium Oxide
Variations: Ceramic Abutment is available only in one form. The total height of the
abutment together with octa is 13mm, whereas the upper diameter is 8mm. It is
indicated for single and multiple unit, cemented restorations where the high level of
aesthetics is required.
Dyna Octalock® Titanium Abutment (Ti)
In order to increase the mechanical strength of the abutment in challenging though
demanding clinical situation, besides the ceramic abutment, Dyna Dental produces the
titanium trimmable abutment. It has the same form as the ceramic abutment and is
ideal for situations where strong and resistant prosthetic constructions are indicated.
Material: Titanium grade 1
Variations: The titanium Abutment is available only in one form. The total height of
the abutment together with octa is 14mm, whereas the upper diameter 11.5mm. It is
indicated for single and multiple unit, cemented restorations.


Impression abutment and implant analogue
The impression abutment is used when transferring the
position of the implant in the mouth to the model. It has the
same octa features as other abutments and provide exact
replication of the octagon position to the working model. It is
intended to be used with open-tray impression technique. For
more information about impression taking please see Dyna
Implant Manual.
Materials: Titanium
Variations: The Impression Abutment is available in one
form. It can be individually adjusted if the interocclusal space
is too small. It is intended to be used for single and multipleunit
restorations and bar retained overdenture restorations.
Dyna Octalock® Healing Abutment (Ti)
Healing abutments are used in the second stage surgery to heal and form the gingiva .
They have special height markings helpful in reading the gingiva height after the
healing period.
Depending on the type of prosthetic construction different healings should be used for
overdentures and crown and bridge works.
Different height of the healing abutments allows better tissue management especially
for fixed constructions.
Materials: Titanium
Variations: The Healing Abutment is available in two basic forms:
 for overdentures
 for fixed constructions
Abutments for fixed constructions are additionally available in high, medium and low
version


1.4.7 Abutments for overdentures.
The abutments for overdenture have only been adjusted to fit the new Octalock
connection. They are almost the same as for the old system. Therefore, please refer to
Dyna Implant Manual for further clinical details.
Materials: bar abutment and ball abutment – titanium, magnet abutment EFM alloy
Variations: all overdentures are available in different heights
-bar - L2=2mm, L3=3mm, L4=4mm, L5=5mm, L6=6mm
-ball - L2=2mm, L3=3mm, L4=4mm, L5=5mm, L6=6mm
-magnet - L3=3mm, L4=4mm, L5=5mm
*depends on the implant diameter see product catalogue
Explanation of the used codes.
AS abutment screw
CP ceramic abutment
HE healing abutment
IA implant analogue
IC impression coping
IS impression screw
MY memory abutment
ST straight abutment
TA temporary abutment
TP titanium abutment
UA universal abutment
UP universal plastic abutment
US universal sleeve
BE bar extension
BL bar abutment
BS bar screw
MG magnet abutment
81 fixed constructions
82 overdentures
The last number in the abutments code means either the abutment
length(2-6), or the presence of the octagon (0-1), exceptionally it
means that the product is only in one design (0)
830 implants 3.6mm
840 implants 4,0mm
The last two numbers in the implant code represents the length of the
implant.